| Structure |
: |
Square tubular Electrostatic power coated steel sections. |
| Top tray |
: |
Ring top tray for monitor.
|
| Table Top |
: |
Stainless Steel. |
| Drawer |
: |
One Large storage drawer. |
| Cylinder Yokes |
: |
Gas specific yokes & gauges with claming bars, dual for Oxygen and Nitrous Oxide. |
| Mobility |
: |
For large diameter anti static castors have brakes. |
| Pressure Gauges |
: |
Two each for Oxygen & Nitrous Oxide. |
| Regulators |
: |
For higher efficiency & reliability two each for oxygen & Nitrous oxide. |
| Non-return cum pressure release valve |
: |
Minimises risk of back flow of gases, blows off when pressure exceeds 200cm H2O. |
| Oxygen failure warning device
|
: |
Pneumatic device, which activates audible alarm (for minimum 10 seconds) when oxygen supply pressure falls to 205 kPa. |
| Vaporisers |
: |
Space for three vaporizers provided as standard Vaporisers are optional. |
| Flowmeters |
: |
Long (up to 260mm) rotating bobbin flow meters calibrated in double/triple scale, antistatic & non-interchangeable.
Oxygen – 100ml/min. to 10 litres/min.
Nitrous oxide –100ml/min to 12 litres/min
Oxygen being a preferred gas kept on downstream of other gas.
|
| Oxygen flush |
: |
Emergency oxygen flush outlet button at table level (36-50lit/min) |
| Change over unit |
: |
Change over unit is provided for open or close circuit. |
| Breathing circuit |
: |
Standard magill's circuit. |
| Nitrous cut-off |
: |
Provide Nitrous cut-off system prevents risk of administering 100% Nitrous oxide in absence of oxygen pressure. |
| Standard accessories |
: |
One/two built – in oxygen outlets (422 kPa) for driving ventilators, (sufficient space for ventilator), extended rear platform for two 10 ltrs water capacity cylinders, key spanner for a type cylinders, facemask for adult and paediatric. |
| Optional |
: |
Calibrated Vaporisers : Halothane / Isoflurane vaporiser are agent specific, colour coded selectatec or cage mount, temperature flow & pressure compensated with calibration (in different models) |
| |
|
Uncalibrated Vaporisers : Goldman / Ether Vaporisers are supplied on demand. |
| |
|
Oxygen ratio controller : To prevent administering less than 25% oxygen in patient circuit. |
| |
|
Circle Absorber : Single / double chamber as required. |
| |
|
Bain Circuit (adult & paed) Extra Rebreathing bag (antistatic) 0.1, 1, 1.5 litre. |
| |
|
Gas Scavenging system : Provides active and Passive waste / expired gas scavenging. |
| |
|
Flowmeters can be configured for 3-tube flowmeter (with air) or 5 tube cascade for low flow. |
| |
|
Basal Flow, Cylinders, Bulck conversion unit for O2 & N2O |
| |
|
Laryngoscopes, Oxygen therapy equipments etc. |
| |
|
Anaesthesia Ventilators. |
| |
|
Anaesthesia Systems are also available in stainless steel structure. |